Analyzer MNT_1.2

Parameters. Scanning range: (length, width, height) 45x45x25 microns. The maximum frame size in pixels, 512 x 512. Standard speed approaching the probe to the surface is 70 nm / ms (achieved with a radius of 50 nm nanocapillary using MNT-z positioner 25 microns). At a sampling rate of 20 kHz and the approach velocity of 100 nm / s vertical resolution up to 5 nm. Lateral resolution is determined by the size of the scan area, the number of pixels and the sample diameter (nanocapillary). Can be used nanocapillaries diameter of 30 nm. Nanocapillaries smaller diameter require less feed rate, resulting in slower scan.   



Lateral resolution is determined by the size of the scan area, the number of pixels and the sample diameter (nanocapillary). Optimal diameter of nanicapillaries 30nm, nanocapillaries smaller diameter require less feed rate, resulting in slower scan.

Approach speed on (nm / ms)

The amplitude of retraction nm )

Time per pixel (msec )

The frame size in pixels

Obtaining time frame (s)

70

700

20,7

32х32

21,2

70

1000

25,3

256х256

1657,1

70

5000

86,4

256х256

5664,2

 